US Technical Specifications

Weight

Refrigeration unit 44 kg
Caps between 795 g and 900 g (incl. coolant and cover)

DIMENSIONS

Height 650 mm x Width 320 mm x Depth 420 mm
Extended height 1650 mm

POWER SUPPLY

100-110 V 50 Hz, 100-120 V 60 Hz or 220-240 V 50/60 Hz single phase 850 VA
All electrical circuits earthed and protected with fused system
All fuses are Bussmann time delay fuses – 10 A at 250 Vac with T-speed

CONDITION OF USE

Temperature: 10-30⁰ C (50-86⁰ F)
Atmospheric pressure: 70-106 kPa
Relative Humidity: 30-75 %

TEMPERATURE CONTROL

Electronic thermostat with touch screen controller
Coolant flow rate and temperature continuously monitored in line

INFORMATION SIGNALS

Visual and audible for high and low temperature conditions
Loss of coolant flow

INLET PRESSURE OF COOLANT FLOW

Pressure into cap – 17 psi
Flow rate of coolant 10-50 ml/s

REFRIGERATION

Hermetically sealed unit. CFC-free R134A refrigerant.

COOLANT

Orbis C – Non-hazardous secondary coolant.
MSDS available upon request.

DEVICE CLASSIFICATION

CE marked in accordance with annex V of the Medical Devices Directive
93/42/EEC for a Class IIa Device. Notified Body No. CE 2797.
MDSAP 699742.

SAFETY STANDARDS

BS EN 60601-1:2006 + IEC 60601-1:2005 for medical electrical equipment,
BS EN 60601-1-2:2015 + IEC 60601-1-2:2014 electromagnetic compatibility

ELECTRICAL CLASSIFICATION

Class 1 type BF

Product specifications are subject to continued evaluation and improvements and subject to change without prior notification. Contact us for more information.
Contact
© Paxman Coolers 2022