Refrigeration unit 44 kg
Caps between 795 g and 900 g (incl. coolant and cover)
Height 650 mm x Width 320 mm x Depth 420 mm
Extended height 1650 mm
220-240 V 50/60 Hz single phase 850 VA
All electrical circuits earthed and protected with fused system
All fuses are Bussmann time delay fuses – 10 A at 250 Vac with T-speed
Temperature: 10-30⁰ C (50-86⁰ F)
Atmospheric pressure: 70-106 kPa
Relative Humidity: 30-75 %
Electronic thermostat with touch screen controller
Coolant flow rate and temperature continuously monitored in line
Visual and audible for high and low temperature conditions
Loss of coolant flow
Pressure into cap – 17 psi
Flow rate of coolant 10-50 ml/s
Hermetically sealed unit. CFC-free R134A refrigerant.
Orbis C – Non-hazardous secondary coolant.
MSDS available upon request.
CE marked in accordance with annex IV of the Medical Device Regulations (MDR) 2017/745 for a Class IIa Device. Notified Body No. CE 2797. MDSAP 699742.
BS EN 60601-1:2006 + IEC 60601-1:2005 for medical electrical equipment,
BS EN 60601-1-2:2015 + IEC 60601-1-2:2014 electromagnetic compatibility
Class 1 type BF